1. Safety Philosophy (Golden Rules)

  1. Pressure is the primary hazard — treat all pressurized components as energized.
  2. Automation reduces risk, but does not replace vigilance.
  3. Never defeat interlocks or alarms.
  4. No maintenance on pressurized systems — ever.

2. Pre-Operation Safety Checks (Before Every Run)

A. System Integrity

  • Verify vessel pressure rating > planned operating pressure
  • Inspect:
    • Seals, O-rings, gaskets
    • Tubing and fittings (no scratches, kinks, corrosion)
    • Filters and screens (clean, seated correctly)

B. Safety Devices

  • Pressure Relief Valves (PRVs):
    • Set below MAWP
    • Not isolated or blocked
  • Burst discs:
    • Correct rating
    • No deformation or corrosion
  • Emergency stop (E-stop): tested and functional

C. Instrumentation

  • Pressure transducers calibrated
  • Temperature sensors functional
  • Flow meters zeroed
  • CO₂ detectors active (audible + visual alarms)

3. Safe Start-Up Protocol

  1. Confirm:
    • Vessel properly closed and torqued
    • Collection vessels secured
  2. Begin slow pressurization
    • Ramp pressure gradually (no step changes)
  3. Monitor:
    • Pressure rise rate
    • Temperature stability
    • Unusual noises or vibrations
  4. Verify stable operating conditions before extraction begins

⚠️ Abort start-up immediately if pressure spikes or temperature drifts abnormally.


4. Safe Operation Protocol

A. During Extraction

  • Maintain operation within validated pressure–temperature window
  • Monitor:
    • Pressure drop across bed
    • CO₂ flow consistency
    • Separator conditions
  • Never:
    • Tighten fittings under pressure
    • Bypass alarms
    • Leave system unattended without alarms enabled

B. Automation & Interlocks

  • PLC/SCADA must:
    • Prevent over-pressure
    • Shut down on sensor failure
    • Enforce correct sequence (pressurize → extract → depressurize)

5. Safe Depressurization Protocol (MOST CRITICAL STEP)

  1. Depressurize slowly
    • Controlled pressure ramp down
  2. Maintain temperature to avoid:
    • Ice formation
    • Seal embrittlement
  3. Vent CO₂ to:
    • Designated exhaust
    • Well-ventilated area
  4. Confirm zero pressure before opening any component

🚫 Never:

  • Rapidly vent manually
  • Open vessels without pressure verification

6. Maintenance & Servicing Safety

Lock-Out / Tag-Out (LOTO)

  • Isolate CO₂ supply
  • Depressurize system fully
  • Lock valves and power sources
  • Attach warning tags

Component Replacement

  • Replace seals on a schedule (not on failure)
  • Use manufacturer-approved parts only
  • Pressure test after reassembly

7. Emergency Safety Protocols

A. CO₂ Leak

  1. Evacuate immediately
  2. Activate emergency ventilation
  3. Do not attempt repair
  4. Re-enter only after CO₂ sensor clearance

B. Over-Pressure Alarm

  • Automatic shutdown
  • Controlled depressurization
  • Root cause investigation required before restart

C. Vessel or Fitting Failure

  • Evacuate
  • Lock out system
  • Incident reporting & inspection

8. PPE Requirements (Last Line of Defense)

PPEUse
Face shieldJet and fragment protection
Safety gogglesEye protection
Cryogenic glovesCold burns
Cut-resistant glovesTubing & fittings
Hearing protectionRapid venting

9. Documentation & GMP Compliance

Required records:

  • Daily pre-operation checklist
  • Calibration certificates
  • Maintenance logs
  • Alarm and interlock testing records
  • Operator training records
  • Incident / deviation reports

Auditors expect:

  • Traceability
  • Preventive maintenance program
  • Emergency drills documentation

10. Safety Inspection Checklist (Summary)

✔ Vessel rating verified
✔ PRVs & burst discs active
✔ CO₂ detectors working
✔ Alarms tested
✔ Ventilation operational
✔ SOP followed
✔ Operator trained


Final Safety Rule

If you are unsure — STOP.
Pressure systems do not forgive improvisation.


Bottom Line

Safe SFE operation depends on engineering controls, disciplined procedures, trained operators, and documentation. Equipment safety protocols must be non-negotiable and auditable, especially under GMP and industrial operation.


If you want, I can:

  • Convert this into a formal SOP
  • Create a one-page operator safety checklist
  • Build a HAZOP-based equipment safety matrix
  • Prepare a training poster or infographic

Just tell me which one you want next.